Mass spectrometry

9 Mass spectrometry





Keypoints








Ion generation


The two most popular methods for ion generation are electrospray ionisation (ESI) and electron impact ionisation (EI). Also popular in certain areas of application is matrix assisted laser desorption ionisation (MALDI).



Electrospray ionisation (ESI)


ESI began to be popularised around 20 years ago and has revolutionised the applicability of MS as an analytical tool. Thus ESI is now the most widely applied method of ionisation because of its ready compatibility with high-pressure liquid chromatography (HPLC). The ionisation takes place under atmospheric pressure. The basis of the technique is shown in Figure 9.1.



The eluent from a HPLC system passes through a quartz or metal needle to which a high electrical potential, e.g. 4.5 kV, is applied. If a positive potential is applied, then the negative ions in the eluent are stripped away by being attracted to the needle thus leaving positively charged solvent droplets which spray out of the capillary. Under the influence of a coaxial flow of nitrogen gas, the droplets evaporate and, as they do so, break up due to internal charge–charge repulsion. In the end gas phase ions are produced which are attracted into the mass spectrometer by an opposite charge applied to a heated capillary which allows a slow bleed from the atmosphere into the mass spectrometer, which has to operate under high vacuum. In order to maintain high vacuum in the instrument two pumping stages are used, an intermediate stage immediately after the heated capillary and a high vacuum stage in the ion separation stage.


Figure 9.2 shows an electrospray spectrum of the basic drug acebutolol, the positive charge on the molecule is generated by its basic centre being protonated, this occurs even in a weakly acidic HPLC mobile phase. The main ion seen at m/z 337.2 (m/z stands for mass over charge ratio) is due to the molecular weight of the drug plus a proton. ESI is known as a soft ionisation technique and in Figure 9.2 it can be seen that there is little fragmentation of the molecule. A very minor fragment can be observed at m/z 319.2, which is due to the loss of water (18 amu). Two additional features of the spectrum should be noted:



1. The ion separation used in this case was based on a quadrupole (discussed later) which measures masses to approximately one decimal place and the mass measured for protonated acebutolol (C18H29N2O4) is greater than the mass of 337 amu, which would be obtained by adding up all the atoms making it up according to their nominal masses, i.e. C=12, H=1, N=14 and O=16. The greatest deviation from nominal mass is for hydrogen (Table 9.1), which has an exact mass of 1.00783. Hydrogen is very abundant in drug molecules and the presence of 29 hydrogen atoms in protonated acebutolol increases its mass by 29 × 0.00783 = 0.23 above its nominal mass of 337 and thus the quadrupole measures it at approximately 337.2 amu.


2. There is an additional peak which is abundant in the mass spectrum at m/z 338.2. This is due to protonated acebutolol where one of the 18 carbon atoms making up the molecule is due to a 13C atom. There is a 1.1% probability that a 13C atom will occur and where there are 18 C atoms there is a 1.1×18 = 19.8% probability that one of them will be a 13C atom. Thus the ion at m/z 338.2 is approximately 20% of the intensity of the ion at m/z 337.2. The abundances of the isotopes for H, N and O are much lower and do not contribute very much in this case. The halogen atoms chlorine and bromine have abundant isotopes and their isotope peaks can be useful in structure elucidation. The exact masses and isotope abundances of atoms commonly found in drug molecules are summarised in Table 9.1.




ESI can also produce negative ions if the polarity on the spray needle is reversed so that it attracts positive ions thus producing a negatively charged spray. Figure 9.3 shows the ESI mass spectrum of ketoprofen, which forms a negatively charged anion at m/z 253.1. In addition, the spectrum is dominated by an adduct ion formed between ketoprofen and formate, which was used as an additive in the HPLC mobile phase. Adduct formation can occur both in positive and negative ion mode but is usually more pronounced in negative ion mode. It can occur even with trace levels of contaminants in the mobile phase, and it can be seen that the spectrum of ketoprofen also contains an ion at m/z 289.1, which is due to adduct formation with traces of Cl in the mobile phase. It can also be observed that there are more background ions present in the spectrum of ketoprofen and the ions at m/z 91.0 and 137.0 are due to clusters of 2 and 3 formic acid molecules. Table 9.2 lists common adduct ions which are observed in ESI spectra.



Table 9.2 Some of the commonly observed additions to molecular ions observe under electrospray ionisation (ESI) conditions





































  Adduct Comment
Addition + ve ion amu
18 NH4+ More likely to occur with ammonium in the mobile phase
22 Na+ Forms readily with traces of Na+ in mobile phase
32 CH3OH With methanol in mobile phase
39 K+ Less common than Na, adduct forms with traces of K+ in the mobile phase
41 CH3CN Formed with acetonitrile in mobile phase
54 CH3OH/Na+ Formed with methanol in mobile phase + traces of Na+
63 CH3CN/Na+ Formed with acetonitrile in mobile phase + traces of Na+
















Addition −ve ion amu
35 Cl Formed with traces of Cl in mobile phase
45 HCOO With formate in mobile phase
60 CH3COO With acetate in mobile phase




Electron impact ionisation (EI)


EI is not compatible with the use of HPLC as a method for introducing the sample into the mass spectrometer. Before ESI was developed several interfaces were developed which were compatible with EI type ionisation, such as particle beam, thermospray and FRIT-EI interfaces, but these have been almost completely superseded by the ESI interface. However, EI is still used in conjunction with sample introduction either via a direct heated probe or via gas chromatography (GC):



Figure 9.5 shows the EI spectrum of ketoprofen. Ions generated under EI conditions are always positive and thus the ion at m/z 254 is exactly the molecular weight of ketoprofen minus the very small mass of an electron. If there is an electronegative atom in the molecule, such as oxygen or nitrogen, usually the positive charge is located there. Unlike the ESI spectrum of ketoprofen shown in Figure 9.3 the EI spectrum of ketoprofen contains many fragment ions because of the high energy nature of the ionisation process. This is an advantage when it comes to structural confirmation since the spectrum provides a unique fingerprint of the molecule which can be matched against a library spectrum and also can be interpreted. Identifying all of the fragment ions in an EI spectrum is often not easy and ketoprofen provides a less common example where many of the fragment ions can be explained (Fig. 9.6).




Figure 9.7 shows a generalised scheme for decomposition of a molecule under EI conditions. The principles of the scheme are as follows:




To summarise, the following rules apply to mass spectrometric fragmentation:



Table 9.3 shows typical fragments which are lost under EI conditions to give the complex fingerprint pattern of a particular molecule. Further examples of EI spectra will be discussed later in the chapter.


Table 9.3 Common losses from a molecular ion



















































































































Loss amu Radicals/neutral fragments lost Interpretation
1 Often a major ion in amines, alcohols and aldehydes
2 H2  
15 CH3· Most readily lost from a quaternary carbon
17 OH· or NH3  
18 H2O Readily lost from secondary or tertiary alcohols
19/20 F·/HF Fluorides
28 CO Ketone or acid
29 C2H5·  
30 CH2O Aromatic methyl ether
31 CH3 Methyl ester/methoxime
31 CH3NH2 Secondary amine
32 CH3OH Methyl ester
33 H2O + CH3·  
35/36 CI·/HCI Chloride
42 CH2= C=O Acetate
43 C3H7· Readily lost if isopropyl group present
43 CH3CO· Methyl ketone
43 CO + CH3·  
44 CO2 Ester
45 CO2 Carboxylic acid
46 C2H5OH Ethyl ester
46 CO + H2O  
57 C4H9·  
59 CH3CONH2 Acetamide
60 CH3COOH Acetate
73 (CH3)3Si· Trimethylsilyl ether
90 (CH3)3SiOH Trimethylsilyl ether


Matrix assisted laser desorption ionisation (MALDI)


MALDI uses a nitrogen laser to promote ionisation of molecules prior to ion separation in a mass spectrometer. It is usually combined with time of flight (TOF) separation of the ions generated. In order for the sample to be ionised it has to be dissolved in a matrix that absorbs UV radiation at around the wavelength (337 nm) produced by the laser. A simple example of a matrix is dihydroxybenzoic acid and there are a number of similar aromatic compounds which are used to promote ionisation of different classes of molecules. The sample solution is mixed with matrix solution on a metal plate and allowed to dry prior to being introduced into the instrument. The laser is then directed at the target plate to promote ionisation (Fig. 9.8).



The technique like ESI is a soft ionisation technique and the ions generated are most commonly due to protonated, or in negative ion deprotonated, molecular ions without extensive fragmentation occurring. It has been widely applied in the characterisation of proteins since it allows ionisation of these high-molecular-weight compounds as singly charged ions and, in combination with TOF separation, measurement of high-molecular-weight species can be carried out. It is also a useful technique for the determination of very polar compounds such as DNA oligomers (DNA fragments 10–20 base pairs long) which contain a phosphate backbone and do not readily ionise in ESI mode because of the strong association of these molecules with sodium ions.


MALDI is also useful for the ionisation of other polar biomolecules and Figure 9.9 shows the ions generated from coenzyme A and two acyl CoAs using MALDITOF in negative ion mode. Another use of MALDI in pharmaceutical development is in the characterisation of pharmaceutical polymers. These usually comprise a mixture of chain lengths and MALDIMS can provide a molecular size distribution for these materials providing one method for setting a quality standard for such materials. The disadvantage of MALDIMS is that it is not readily quantitative and cannot be linked directly with HPLC. However, another area where MALDI has proved useful has been in tissue imaging where a tissue section is coated with matrix and the MALDIMS instrument is used to scan across the tissue in ca 0.1 mm steps to image the tissue in terms of the molecules making it up which could for instance lead to a better understanding of a disease process or find the site of accumulation of a drug within the tissue.



Figure 9.10 shows a MALDIMS image of an experimental antipsychotic drug in rat brain, one of the major sites of accumulation is in the cerebellum.


< div class='tao-gold-member'>

Stay updated, free articles. Join our Telegram channel

Jun 24, 2016 | Posted by in PHARMACY | Comments Off on Mass spectrometry

Full access? Get Clinical Tree

Get Clinical Tree app for offline access