Simultaneous Quantitation of Estradiol and Estrone in Serum Using Liquid Chromatography Mass Spectrometry


Calibrator number

Volume of previous calibration material (ml)

Volume of deionized water (ml)

Final concentration (pg/ml)

6

3.75 (working stock)

2.25

625

5

2.0 (Cal #6)

4.0

208

4

2.0 (Cal #5)

4.0

69

3

2.0 (Cal #4)

4.0

23

2

2.0 (Cal #3)

4.0

7.7

1

2.0 (Cal #2)

4.0

2.6




 


2.

Controls: Bio-Rad Lyphochek Immunoassay Plus Control Level 1–3: To prepare controls 2–4 add appropriate volume of working estrone control material and deionized water to each vial of lyphochek for a total volume of 5.0 ml (see below, Table 2.) and allow to stand for 15 min prior to use. Mix by swirling. Control 1 will be prepared by diluting control 2 1:10 with deionized water during each run. Aliquot into 300 μl aliquots and store at <−20 °C. Stable for up to 30 days (see Note 3 ).


Table 2
Preparation of controls





























Control

Volume of control estrone material (μl)

Volume of deionized water (ml)

Final concentration of estrone (pg/ml)

2

500

4.5

100

3

1000

4.0

200

4

2000

3.0

400

 

3.

Operating parameters: Set instrument parameters according to Tables 3 and 4. Parameters are optimized specifically for an API 5500 with Shimadzu 20A Prominence HPLC System; therefore, tune setting may vary slightly between instruments.


Table 3
HPLC operating conditions

































































Column temp

45 °C
 

Flow rate

0.45 ml/min
 

Equilibration time

0.5 min
 

Gradient

Step

Time (min)

Parameter (%)
 
1

0.3

45
 
2

0.5

45
 
3

2.00

80
 
4

2.20

80
 
5

2.50

90
 
6

2.80

90
 
7

3.00

45
 
8

3.50

Stop



Table 4
Mass spectrometry operating conditions

































Mode

MRM

Polarity

Negative

Curtain gas

15.0 psi

Ion source gas 1

70.0 psi

Ion source gas 2

75.0 psi

Nebulizing current

−5.00

Collision cell exit potential

−15.00

Entrance potential

−13.0

TEMP

700.0

 





2.5 Equipment and Supplies




1.

2.0 ml 9 mm Short-cap Screw Thread Vial (Restek, Bellefonte, PA).

 

2.

2.0 ml 9 mm Short-cap Screw Thread Vial Closure (Restek, Bellefonte, PA).

 

3.

Vial Inserts, 100 μl, tri-spring (Phenomenex, Foster City CA).

 

4.

13 × 100 ml Aliquot Tube.

 

5.

Eppendorf Pipette Tips EP 2–200 μl.

 

6.

Eppendorf Tips, 2.5 ml.

 

7.

Repeater pipet tips 50 ml.

 

8.

Centrifuge: Capable of Speeds of 3267 × g (RCF).

 

9.

TurboVap® IV Evaporator (Zymark Corporation, Hopkinton, MA).

 

10.

Instrumentation: API 5500 with Shimadzu 20A Prominence HPLC System (AB Sciex, Foster City CA).

 

11.

Software: Analyst 1.6 quantitative software and Multiquant Software Version 2.1 (AB Sciex, Foster City CA).

 

12.

Analytical Column: Kinetex XB-C18, 2.6 μl, 50 × 2.10 mm 100 Å (Phenomenex, Torrance, California).

 



3 Methods



3.1 Stepwise Procedure




1.

Label an appropriate number of 13 × 100 ml glass tubes: Cal 1–6, Blank (Negative/Carry-Over), QC1, QC2, QC3, & QC4.

 

Oct 21, 2016 | Posted by in BIOCHEMISTRY | Comments Off on Simultaneous Quantitation of Estradiol and Estrone in Serum Using Liquid Chromatography Mass Spectrometry