Quantitation of Ubiquinone (Coenzyme Q10) in Serum/Plasma Using Liquid Chromatography Electrospray Tandem Mass Spectrometry (ESI-LC-MS/MS)


Calibrator

Volume of stock standard (μL)

Volume of ethanol (μL)

Final concentration (μg/mL)

Final dilution

6

200


6.0


5

200

200

3.0

1:2

4

333

667

2.0

1:3

3


667

0.67

1:3

2


667

0.22

1:3

1


667

0.07

1:3




 


2.

Controls: Chromsystems coenzyme Q10 plasma control Bi-Level I+II (Chromsystems, GmbH, Gräfelfing, Germany): Control material is prepared per the manufacturer’s instructions; 2.0 mL of deionized water is added to the lyophilized material and allowed to stand at least 15 min prior to use. A negative control is prepared utilizing 3.0 % bovine albumin .

 





2.5 Equipment and Supplies




1.

Instrumentation: Applied Biosystems Inc. API 3200 with Shimadzu Prominence 20A HPLC system equipped with Analyst software version 1.6.1. Operating parameters are described in Table 2a, b. Parameters have been optimized specifically for the API 3200 with Shimadzu 20A Prominence HPLC System; some variations would be expected as tune settings often vary slightly between instruments.


Table 2
ESI-LC-MS/MS operating conditions






























































































































































A. HPLC

Column temp

40 °C
 

Flow rate

0.80 mL/min
 

Gradient

Step

Time (min)

Module

Event

Parameter (%)
 
1

0.10

Pumps

Pump “B” conc

0.0
 
2

0.50

Pumps

Pump “B” conc

25
 
3

1.00

Pumps

Pump “B” conc

60
 
4

1.10

Pumps

Pump “B” conc

60
 
5

2.00

Pumps

Pump “B” conc

80
 
6

2.10

Pumps

Total flow

0.85 mL/min
 
7

2.10

Pumps

Pump “B” conc

80
 
8

2.80

Pumps

Pump “B” conc

100
 
9

3.50

Pumps

Pump “B” conc

100
 
10

3.70

Pumps

Pump “B” conc

0.0
 
11

5.00

System controller

Stop
 

B. MS/MS settings

Scan
 
MRM
   

Polarity
 
Positive
   

Curtain gas
 
20.0 psi
   

GS1
 
30.0 psi
   

GS2
 
35.0 psi
   

NC
 
3.00
   

CXP
 
7.00
   

EP
 
10.0
   

TEMP
 
400.0