Quantitation of S-Adenosylmethionine and S-Adenosylhomocysteine in Plasma Using Liquid Chromatography-Electrospray Tandem Mass Spectrometry



Fig. 1
HPLC -ESI-MS/MS ion chromatogram of SAM (m/z 399.1 > 250.1), SAH (m/z 385.1 > 136.1), SAM-D3 (m/z 402.1 > 250.1) and SAH-D4 (m/z 389.1 > 138.1). Concentration of SAM and SAH is 95 nM and 41 nM, respectively




 






3.2 Data Analysis




1.

Instrumental operating parameters are given in Tables 1 and 2.


Table 1
HPLC -ESI-MS/MS operating conditions































































(a) HPLC (SAM and SAH)a

Column temp.

40 °C
 

Flow rate

0.20 mL/min
 

Gradient

Time (min)

Mobile phase A (%)
 
0

75
 
7

0
 
7.1

75
 
10

Stop

(b) MS/MS tune settingsb

Entrance Potential (V)

10

Curtain gas (psi)

20

CAD gas

Medium

Ion spray (V)

5000

Temp (°C)

700

GS 1 (psi)

40

GS 2 (psi)

50

Resolution Q1 and Q3

Unit


aOptimized for Shimadzu Nexera liquid chromatography system equipped with Phenomenex EZ:faast, 3 μm, 250 × 2 mm analytical column; Mobile phase A: 4 mM ammonium acetate, 0.1 % formic acid and 0.1 % heptafluorobutyric acid in water; Mobile phase B: 0.1 % formic acid in methanol

bOptimized for AB Sciex 5500QTRAP®. Tune settings may vary slightly between instruments



Table 2
HPLC -ESI-MS/MS operating conditions















 
MRM transition

Compound