Fig. 1
HPLC -ESI-MS/MS ion chromatogram of 5-MTHF 100 nM standard. [5-MTHF 1 (m/z 460.3 > 313.3), 5-MTHF 2 (m/z 460.3 > 194.5), and 5-MTHF-13C5 (m/z 465.3 > 313.3)]
Fig. 2
HPLC -ESI-MS/MS ion chromatogram of 5-MTHF abnormal QC (30 nM). [5-MTHF 1 (m/z 460.3 > 313.3), 5-MTHF 2 (m/z 460.3 > 194.5), and 5-MTHF-13C5 (m/z 465.3 > 313.3)]
3.2 Data Analysis
1.
Instrumental operating parameters are given in Table 1.
Table 1
HPLC -ESI-MS/MS operating conditions
(a) HPLC (5-MTHF)a | ||
Column temp | 40 °C | |
Flow rate | 0.375 mL/min | |
Gradient | Time (min) | Mobile phase A (%) |
0 | 100 | |
1.5 | 0 | |
2 | 0 | |
2.1 | 100 | |
(b) MS/MS tune settingsb | ||
Entrance potential (V) | 10 | |
Curtain gas (psi) | 20 | |
CAD gas | Medium | |
Ion spray (V) | 5500 | |
Temp (°C) | 500 | |
GS 1 (psi) | 50 | |
GS 2 (psi) | 50 | |
Resolution Q1 and Q3 | Unit |
2.
Data are analyzed using Analyst software version 1.6.2 (AB Sciex).