CHAPTER 216 Temporary Pacing
External (Transcutaneous) Pacing
Indications
• Short-term pacing until transvenous pacing can be initiated or underlying conditions requiring pacing are corrected (e.g., drug overdose, hyperkalemia)
• When medical therapy is not immediately available, or when significant bradyarrhythmias have not responded to medical therapy (e.g., atropine, isoproterenol)
• Symptomatic patients (e.g., syncope, presyncope, dizziness, fatigue) with type I or type II second-degree AV block, third-degree or complete AV block, asystolic pauses exceeding 3 seconds, or an escape pacemaker rate less than 40 beats per minute (bpm); also for patients who become symptomatic because of a bifascicular block or sinus node dysfunction
• As a standby or prophylaxis before surgery in patients with a preexisting cardiac conduction block (anesthesia can exacerbate the block); also before cardiac diagnostic studies
• As a standby or prophylaxis in conscious patients with an expected bradyarrhythmia or a new type II second-degree, third-degree, or complete AV heart block in the setting of ischemia or infarction (frequently seen with acute anterior or inferior wall myocardial infarctions or digoxin overdose)
Contraindications
All of the contraindications are relative.
• Bradycardia in patient with significant hypothermia; as the core temperature drops, the ventricles become more irritable and prone to fibrillation that is resistant to defibrillation. In addition, bradycardia may be physiologic due to a decreased metabolic rate in these individuals.
• Bradycardia in children: usually due to hypoxia or hypoventilation, the best intervention is to provide an adequate airway as opposed to pacing (exceptions as mentioned in the Indications section).
• Overdrive pacing in tachyarrhythmias with rates greater than 180 bpm because that is the maximal rate of most external pacers.
• Bradyasystolic or asystolic arrest of more than 20 minutes’ duration because of the well-documented poor resuscitation rates.
• Patient is unable to cooperate or tolerate the procedure (in life-threatening situation, provide sedation).
Equipment
• Two 8-cm electrodes. These are usually round or rectangular and packaged in pairs. The negative electrode may also be labeled “front,” “apex,” or “anterior”; the positive electrode may be labeled “back” or “posterior.”
• Pacing unit (contains pulse generator and monitor). The best units allow either fixed-rate or demand mode. Most allow a range from 30 to 180 bpm, with current output from 0 to 200 mA. Pulse durations vary from 20 to 40 msec and are not adjustable by the operator. To protect health care providers, some pacers shut off when an electrode falls off the chest.
Technique
1 Attach the exposed adhesive surfaces of two large electrode patches to the anterior and posterior chest walls (Fig. 216-1). The negative (anterior) electrode should be placed over the apex (at the point of maximal impulse) or the septum of the heart (the lead V3 position), and the positive (posterior) electrode directly behind the anterior electrode, to the left of the thoracic spine, between the spine and the left scapula. Alternatively, the positive electrode can be placed on the right upper chest with the negative electrode over the apex of the heart.
2 If time allows, prepare the patient with analgesia such as a narcotic or sedation with a benzodiazepine (especially helpful if the required mean current for capture is ≥50 mA).
3 Turn on the pacing unit. Set the heart rate (e.g., 80 bpm) to demand pacing. Set the current output and sensing thresholds at levels similar to those used for internal pacing; however, remember that larger outputs are necessary. Keep in mind that for demand pacing, external units often do not have sensing thresholds. The final current output setting is usually 1.25 times the initial capture threshold. Patients with conditions that cause difficult or ineffective pacing may require higher outputs for capture. At these higher outputs, the resultant muscle twitches may be so severe as to preclude external pacer use.
4 Apply the electrical stimulation to the electrodes. For conscious patients, slowly increase the output from the minimal setting, at 5- to 10-mA increments, until capture is achieved. Electrical capture is usually indicated by a widening of the QRS complex and especially by a broad T wave. The output required to obtain capture is defined as the pacing threshold. For asystolic or unconscious patients, begin at full output (200 mA) and decrease until capture is achieved, which defines the pacing threshold. The final current output should be set at the pacing threshold or 5 to 10 mA above it. When transcutaneous pacing is used as a standby technique, most clinicians document capture by initiating a brief period of pacing at a rate slightly higher than the patient’s intrinsic rate. The pacing output threshold is then recorded and the pacing unit is returned to standby mode.
5 When successful pacing is achieved, prophylactic intravenous access (central venous [CV] catheter) through the right internal jugular vein may be helpful in case an internal pacer is needed urgently. Because external pacers have up to a 20% failure rate, having CV access available minimizes the risk of having to obtain it during a “code” situation.
6 Monitor continuously for capture and potential complications (e.g., treatable ventricular fibrillation, burns). The only sure sign of electrical capture is the presence of a consistent ST segment and T wave after each pacer spike. Palpation of the carotid to confirm a pulse is not helpful because the muscle stimulation and contractions produced by the pacer simulate a carotid pulse.
Internal (Transvenous) Pacing
Indications
Therapeutic
• Symptomatic, hemodynamically compromising or life-threatening bradyarrhythmias unresponsive to pharmacologic therapy (e.g., systolic blood pressure <80 mm Hg, change in mental status, angina, pulmonary edema), including sick sinus syndrome
Contraindications
Contraindications include those listed in Chapter 211, Central Venous Catheter Insertion, and those listed previously for external pacing. Other contraindications include the following:
• A situation in which the bradycardia is well tolerated and the symptoms are intermittent, mild, or rare.*
• Bradycardia in patients with significant hypothermia; as the core temperature drops, the ventricles become more irritable and prone to fibrillation (especially if the pacing wire contacts the heart muscle) that is resistant to defibrillation. In addition, bradycardia may be physiologic because of a decreased metabolic rate in these individuals (relative contraindication).
• Digoxin toxicity and other drug ingestions that may increase the irritability of the myocardium (relative contraindication in life-threatening situation).
• Presence of a prosthetic tricuspid valve (relative contraindication in life-threatening situation).
• Depending on access site, planned neck or clavicle surgical procedures (relative contraindication, may affect choice of site).
• Distortion of local anatomy or landmarks; for insertion from subclavian, moderate to severe chest wall deformities that distort local anatomy (relative contraindication, may affect choice of site).
• Suspected injury to the superior vena cava (relative contraindication, may affect choice of site; e.g., superior vena cava syndrome, in which insertion from below the diaphragm is preferable).